R., Widiatmono and J., Mandon and F.J.M., Harren and Kusminarto, Kusminarto and M.A.J., Wasono and Mitrayana, Mitrayana (2014) QCL BASED INTEGRATED CAVITY OUTPUT SPECTROSCOPY FOR CO GAS DETECTION. Proceeding of International Conference On Research, Implementation And Education Of Mathematics And Sciences 2014. (Submitted)

[img] Text
09-Restu Widiatmono.rtf

Download (1MB)
Official URL: http://fmipa.uny.ac.id

Abstract

A sensitive gas detection system using a distributed feedback CW Quantum Cascade Laser (QCL) centered at 4610 nm as a radiation source has been developed. QCL was used in combination with an off-axis integrated cavity output spectroscopy for Carbon Monoxide (CO) gas detection in a low gas pressure. a 15 cm long high-finesse cavity with an effective optical path length of 400 m has been used as a sample gas cell. We demonstrated detection limits of an up to 1 ppb level with less than 2 second averaging time. The result clearly showed that The QCL based ICOS system is a robust technique that suitable for fast and sensitive gas detection.